Preparing the System
To remove the tray
Gently lift the tray at the front and then pull it back to slide it out of the guides.
Setting up the System for Analyses Using the Flame Technique
Summary of the procedure
Switching on the System on page 165
2. If you need to change techniques from furnace technique to the flame
technique, go into WinLab32 for AA and select
File > Change Technique >
flame
.
3. Create or open a method (see the
Online Help in WinLab32 for AA
).
4. If you intend to make emission measurements, in the
Method Editor
on the
Inst
page, for
Signal Type
, select
Emisssion
.
5. If you intent to make absorption measurements, install the lamps that you
Installing the Lamps on page 168
6. Set up the burner system (see
Setting Up the Burner System on page 176
)
7. Perform the safety checks (see
Safety Checks for the Flame Technique on page
8. Ignite the flame (see
Igniting the Flame on page 178
)
You should now set up the nebulizer, then optimize the burner and nebulizer before
checking the sensitivity. Refer to the
Online Help in WinLab32 for AA
.
Setting Up the Burner System
Summary of the procedure
1. Make sure that the correct nebulizer is installed – the stainless steel nebulizer
or a high sensitivity plastic nebulizer (see
Nebulizer Installation on page 132
).
Summary of Contents for PINAACLE 900 Series
Page 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Page 2: ......
Page 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Page 52: ...36 Safety Practices...
Page 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Page 84: ...68 Preparing Your Laboratory...
Page 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Page 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Page 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Page 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Page 210: ...194 Preparing the System...
Page 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Page 354: ...338 Maintenance...
Page 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Page 389: ......