132
Installation
3. Disconnect the drain tube from end cap.
4. Pour about 250 mL deionized water slowly through the drain tube to fill the
drain trap and activate the drain float.
5. Reattach the drain tube to the end cap.
6. Tighten the clamp around the drain tube.
7. Reinstall the burner head.
8. Close the compartment door.
Nebulizer Installation
The instrument is shipped with the nebulizer already installed. Depending on the
instrument model ordered, it will be shipped with either a standard stainless steel
or a high sensitivity plastic nebulizer.
For information on removing and reinstalling the stainless steel or high sensitivity
plastic nebulizer see
Nebulizer Removal and Reinstallation on page 230.
Installing AS 900 Furnace Autosampler
The AS 900 autosampler can be installed on the PinAAcle 900H, 900T and 900Z.
This autosampler will initially be installed by your PerkinElmer Service engineer
but these instructions are also provided in case you need to alter any of the
connections or move the autosampler in your laboratory.
Note
The figures shown in this section are of a PinAAcle 900T but the
installation of a AS 900 Furnace Autosampler is the same for the PinAAcle
900H and 900Z.
Summary of Contents for PINAACLE 900 Series
Page 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Page 2: ......
Page 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Page 52: ...36 Safety Practices...
Page 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Page 84: ...68 Preparing Your Laboratory...
Page 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Page 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Page 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Page 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Page 210: ...194 Preparing the System...
Page 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Page 354: ...338 Maintenance...
Page 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Page 389: ......