![otometrics Hortmann Vestlab 100 User Manual Download Page 74](http://html.mh-extra.com/html/otometrics/hortmann-vestlab-100/hortmann-vestlab-100_user-manual_1658650074.webp)
72
Otometrics
Test Battery
Calorics test
As we can see, the calculation of the VP has not changed when the velocity of the
spontaneous nystagmus has been subtracted, but the amount of the DP gets signifi-
cantly smaller.
4.7.8.3
Conclusion
To calculate the VP and DP correctly, the results of the irrigations have to be entered
with the correct sign into the sign adapted formulas. Also, the value of a spontaneous
nystagmus has to be subtracted, which doesn’t change the result for the VP, but
changes the DP significantly.
4.7.9
Calorics test editing
The result of the Calorics test is displayed in the test window.
To edit the result double-click on one of the diagrams, or select
Measurement >
Calorics
, or right-click and select
Edit
in the diagram.
.
Directional Preponderance (DP)
Calculation Consideration of Spontaneous Nystagmus
%
100
)
SPN
CL
(
)
SPN
WL
(
)
SPN
CR
(
)
SPN
WR
(
)
SPN
WL
(
)
SPN
CR
(
)
SPN
CL
(
)
SPN
WR
(
%
100
SPN
CL
SPN
WL
SPN
CR
SPN
WR
SPN
WL
SPN
CR
SPN
CL
SPN
WR
%
100
CL
WL
CR
WR
SPN
4
WL
CR
CL
WR
Summary of Contents for Hortmann Vestlab 100
Page 1: ...Hortmann Vestlab 100 User Manual Doc no 7 50 0940_02 Part no 7 50 09400 0459 ...
Page 6: ...4 Otometrics Table of Contents Hortmann Vestlab 100 User Manual ...
Page 16: ...14 Otometrics Introduction Typographical conventions ...
Page 92: ...90 Otometrics Troubleshooting Vestlab Software ...
Page 100: ...98 Otometrics Technical Specifications Hortmann Vestlab 100 ...