Test Battery
Calorics test
Note ·
The test settings include both Position and Positioning tests. Only Positioning test
will be explained here.
The settings can be accessed from the menu bar:
Settings > Tests > Position/Posi-
tioning
.
4.7
Calorics test
4.7.1
Description
Caloric tests provide an objective evaluation of the function of the vestibular laby-
rinths.
Irrigation with either water or air induces nystagmus in patients with a functioning
balance system. The nystagmus resulting from the irrigations in both ears are com-
pared for evaluation.
Position
Duration (s)
Set the test duration of the measurement.
Analysis profile
Select the analysis profile to be used to detect nys-
tagmus in the measurement. The analysis profiles
can be configured from the menu bar:
Settings >
Nystagmus analysis
.
View/Layout
Channel distance (°)
Set the distance between the horizontal and vertical
channel.
Channel scaling (°)
Set the amplitudes displayed for each measure-
ment; 40 will be displayed as -20
°
to +20
°
on the y-
axis.
Hallpike-Stenger
Includes nine re-positionings of the patient during
the test. Follow clinic/hospital procedures.
Dix-Hallpike
Includes four re-positionings of the patient during
the test. Follow clinic/hospital procedures.
Horizontal vertical
Select the channels to be displayed on the screen.
Summary of Contents for Hortmann Vestlab 100
Page 1: ...Hortmann Vestlab 100 User Manual Doc no 7 50 0940_02 Part no 7 50 09400 0459 ...
Page 6: ...4 Otometrics Table of Contents Hortmann Vestlab 100 User Manual ...
Page 16: ...14 Otometrics Introduction Typographical conventions ...
Page 92: ...90 Otometrics Troubleshooting Vestlab Software ...
Page 100: ...98 Otometrics Technical Specifications Hortmann Vestlab 100 ...