When specifying XX as the unit number, set <parameter̲2> to 0.
・ Set the following numerical values for <parameter̲1>.
Once continuous or synchronous measurement has been
specified, the setting will remain in effect until a stop
measurement command is executed. If <parameter̲1> has been
set to 0, <parameter̲2> should also bet set to 0.
・ Set the following numerical values for <parameter̲2>
<parameter̲2> can be omitted. If omitted, the value defaults to 0.
Output Response 1
@<unit̲number>ME̲OK[CR] :Measurement results are not output next
@<unit̲number>ME̲READY[CR] :Measurement results are output next
@<unit̲number>ME̲ER[CR] :Measurement is not possible
because unregistered; incorrect argument
・ <unit̲number> :Unit number (00 to 31)
Output Response 2
When returning results is specified in <parameter̲2>, the results will be
returned continuously in the following format after the READY response
until measurement stops.
47
・ Set the following numerical values for <parameter>.
<parameter> can be omitted. If omitted, the value defaults to 1
Output Response
@<unit̲number>MD̲<comparison̲result><contrast><average̲density>[CR]
: Plain measurement
@<unit̲number>MD̲<comparison̲result><agreement>[CR]
: Pattern measurement at 1 model
@<unit̲number>MD̲<comparison̲result><agreement><agreement>[CR]
: Pattern measurement at 2 models(Wide-mode)
@<unit̲number>MD̲<comparison̲result><agreement><agreement><agreement>[CR]
: Pattern measurement at 3 models(Wide-mode)
@<unit̲number>MD̲ER[CR] : Not executed correctly
<unit̲number>
Unit number (00 to 31)
<comparison̲result> 0: Comparison result was OK
1: Comparison result was NG
<contrast>
0 to 8:Contrast level
<average̲density>
0 to 8:Level of difference between
average density and reference value
<agreement>
0 to 8:Agreement level
Example
To acquire the previous measurement performed by Unit 15 (where Unit
15 is in Normal Mode and is performing pattern measurements):
Input: @15MD[CR]
Output: @15MD̲07[CR]
----------------------------------------------------------------------------------------------
■ ME -MEasure- : Perform measurement and acquire result
----------------------------------------------------------------------------------------------
ME starts and stops measurements. The results are output to the
external signal line in all measurement modes, regardless of the setting
of <parameter̲2>.
Input Command
@<unit̲number>ME̲<parameter̲1>[CR]
@<unit̲number>ME̲<parameter̲1>̲<parameter̲2>[CR]
・ Set the value of <unit̲number> to a number between 00 and 31, or XX
Parameter 1
0
1
2
3
4
Content
Stop measurement.
Start continuous measurement.
Start synchronous measurement.
(Actual measurement is performed when TRG input line turns ON.)
Perform image capture measurement once only.
Perform measurement of image in image memory once only.
Parameter 2
0
1
Content
Results are not returned.
Results are returned as a level (0 to 8).
Unit No.
00 to 31
XX
Content
Unit number of Unit performing measurement.
When all F10's are forced to perform measurements.
Parameter
1
Content
Results are returned as a level (0 to 8).