17.7 Sweep Oscillation Mode
WF1967/WF1968
17-15
Common regardless of sweep type.
However, logarithmic can use only frequency sweep.
Sweep range setting
Specify starting value and stop value, or center value and
span value
However, the center value is also a simple average of
starting value and stop value during frequency logarithmic
sweep.
Assigning a marker value to a center value is possible (the
other way around is also possible).
Setting range of Sweep Time
0.1ms to 10,000s (resolution 4 digits or 0.1ms)
Common regardless of sweep type.
Sweep mode
Continuous, single, gated single (switch)
Common regardless of sweep type.
For gated single, it oscillates only during running sweep.
However, when the waveform is DC, gated single is not
available.
Operation
Start, stop, hold, resume, starting value output and stop
value output.
Trigger source (used for single sweep and gated single sweep)
Internal, external input terminal (switch)
In addition, CH2 can select the same trigger source as
CH1.
Common regardless of sweep type.
Trigger delay setting is disabled.
Manual trigger available.
Internal trigger oscillator for sweep (used for single sweep and gated single sweep)
Period setting range
100.0µs to 10,000s (resolution 5 digits or 0.1µs)
Stop level setting (used for gated single sweep)
Function
Specifies the signal level when gated single-sweep is
stopped.
Setting range
-100.00% to +100.00% (amplitude full-scale reference
and resolution 0.01%) or off
When stop level is set to off, stops by set oscillation
start/stop phase.
Common regardless of sweep type
Remarks
When the waveform is noise, oscillation start/stop phase is
invalid, stop level is always valid.
When the waveform is DC, oscillation start/stop phase is
invalid, stop level is invalid
For phase sweep, sweep starting value is oscillation start
phase and sweep stop value is oscillation stop phase
Oscillation stop unit when gated single Cycle, Half Cycle (switch)
Summary of Contents for WAVE FACTORY WF1967
Page 2: ......
Page 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Page 4: ......
Page 10: ...WF1967 WF1968 vi...
Page 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Page 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Page 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Page 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Page 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Page 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Page 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Page 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Page 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Page 326: ......
Page 328: ......
Page 329: ......