4.4 Setting for Main Items
WF1967/WF1968
4-38
4.4.11 To Set Waveform Polarity and Amplitude Range
a) Setting procedure
1.
Selecting the Polarity/Amplitude Range icon to
the right of the waveform name display will
show the current polarity and amplitude range
settings.
Press the ENTER key to open the selection list
for the polarity/amplitude range.
2.
Select the desired polarity and amplitude range
from the selection list and press the ENTER
key to reflect it in the output.
You can also enter the number of the desired
polarity and amplitude range displayed in the
list on the numeric keypad to reflect it in the
output.
b) What are polarity and amplitude range
You can reverse the polarity or change the amplitude range to single-polarity for each waveform.
The following figure shows the cases with sine waves:
"Reverse" only reverses a waveform without changing the sign of output DC offset.
Check
The polarity and amplitude range setting are separate settings for
each waveform.
Polarity
Amplitude
Normal
Inverted
Selection list (soft-key)
Summary of Contents for WAVE FACTORY WF1967
Page 2: ......
Page 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Page 4: ......
Page 10: ...WF1967 WF1968 vi...
Page 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Page 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Page 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Page 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Page 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Page 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Page 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Page 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Page 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Page 326: ......
Page 328: ......
Page 329: ......