12.3 Defining User-defined Units
WF1967/WF1968
12-3
In the Utility screen, select
the [User Unit] and press
the ENTER key
12.3 Defining User-defined Units
■
Setting screen
The setting is done on the Utility screen.
1.
In the Utility screen, select the [User Unit] field and
press the ENTER key.
2.
The user-defined unit setting window opens. Select each item, and then press the ENTER key
to set the user-defined unit. Alternatively, you can open the input field by pressing the
corresponding soft-key.
The following items can be set in this window:
Setting target [Type]
Select one from among frequency, period, amplitude, DC offset, phase, and duty.
Unit name [Unit]
Sets the unit name with up to 4 characters. Enter a character one by one at the cursor position using
the up/down arrow key, the modify knob, or the numeric keypad.
Formula [Form]
Select either one of [(h+n)*m] or [(log(h)+n)*m] as the unit conversion formula. Where, h is an
original value of the setting target. The value converted by this formula is displayed on the screen
when the user-defined unit is actually used.
The unit of h is defined as shown in the following table depending on the setting target:
Setting target
Unit of h
Frequency
Hz
Period
sec
Amplitude
If amplitude range is
±FS, then Vp-p
If amplitude range is
0/+FS, -FS/0, then Vpk
A value under the designated load impedance condition in each case
DC Offset
V
A value under the designated load impedance condition
Phase
deg
Duty
%
The user-defined unit setting
window opens
Setting target
Unit name
Formula
Multiplier of formula
Offset of formula
Summary of Contents for WAVE FACTORY WF1967
Page 2: ......
Page 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Page 4: ......
Page 10: ...WF1967 WF1968 vi...
Page 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Page 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Page 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Page 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Page 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Page 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Page 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Page 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Page 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Page 326: ......
Page 328: ......
Page 329: ......