6.2 Meaning of Each Parameter and Waveform Examples
WF1967/WF1968
6-11
b) Off-Ph Ctrl Sine
Overview
Sine wave which follows a slope during shut off.
Application examples
•Simulates the output waveform of an AC power supply with a restricted rising and falling time
Meaning of each parameter
The example shows the case where the amplitude range is ±FS.
•Shut off start phase (OffPhase)
The original sine wave ends at this phase.
Variable range: 0.00
°
to 360.00
°
•Shut off slope time (SlopeT)
The linear change in time of the amplitude from the
amplitude of the shut down start phase to the zero.
Variable range: 0.00 % to 50.00 % of basic cycle
The amplitude of the original Sine wave is fixed at ±FS.
Please note that the amplitude may be less than ± FS
depending on the value of the shut down start phase.
Waveform examples
The polarity and amplitude range are all Normal, ±FS.
Summary of Contents for WAVE FACTORY WF1967
Page 2: ......
Page 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Page 4: ......
Page 10: ...WF1967 WF1968 vi...
Page 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Page 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Page 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Page 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Page 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Page 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Page 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Page 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Page 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Page 326: ......
Page 328: ......
Page 329: ......