Chapter 2
Menu Configuration
2-42
MS-850 Raster Scanner User’s Manual
2–Menu Conf
ig
.
Sweep Rate
Default:
10
Options:
0 to 30
Sweep Rate is the number of sweeps per second the raster mirror completes. A sweep
is one pass of the raster, up or down. The maximum sweep rate is limited by the size
of the arc the raster must move through. This table shows the maximum speeds at
selected raster sweep angles.
To maximize the number of scans per label, use the lowest effective sweep rate
required for the application.
Read Cycle On/Off
Default:
Disabled
Options:
Disabled, Enabled
Note:
Read Cycle On/Off will have no effect unless Laser On/Off is also enabled.
When both Read Cycle On/Off and Laser On/Off are enabled, at the end of a triggered
read cycle, the mirror will park at the Top Offset position.
Note:
This feature should only be used when the raster sweep rate is at least twice
the required trigger rate (decodes and outputs per second).
Raster Sweep
Angles
Maximum Sweeps
Per Second
0 to 10°
30
11 to 20°
20
21° to 36°
10