Page 62
Maintenance, Service, Spare Parts Lists
thicknessCONTROL MTS 8202.LLT / 4350127.347
7.3
Software Error Messages
Error message
Actions
Sensor arm is not in upper position
-
Check free movement
-
Check emergency off
-
Wait for active movement
-
Check light barrier (alignment, free path of beam)
Sensor arm is not in lower position
-
Check free movement
-
Check emergency off
-
Wait for active movement
-
Check light barrier (alignment, free path of beam)
Traversing unit is not in home position
-
Check free movement of the carriages
-
Wait for active movement of the unit
Calibration part is not in position
-
Check free movement of pivoting arm
-
Check emergency off
-
Check light barrier (alignment, free path of beam)
-
Check compressed air
Sensor signal very noisy !!! - Please
check and clean the sensors
-
Check ambient light
-
Check for contamination of the sensor glasses on inside
and outside
Waiting for the release of the travers-
ing unit.
-
Wait for time x until the sensor head is in position
Displacement of both axes - Please check
their positions. Accuracy is probably
out of limit!
-
Check distances of carriages to limit switches in homing
position
-
Check marks on the limit switches for misalignment
Test travel canceled - sensor gap is not
empty!!!
-
Check roller for contamination
-
Check measuring gap or roller for material
-
Stop button in software was pressed
The difference between current and de-
livery gap value is too high!
-
Check calibration part and sensors for contamination
-
Check position of laser line on calibration part (cen-
tered?)
-
If ambient temperature is cold: Let sensors warm up
Motor of calibration part does not ro-
tate - please check
-
Check supply voltage according to E plan
-
Check for free movement
-
Check for running noise
-
Check for mechanical fastening
Sensors at risk. Measurement stopped
-
Check light barrier (alignment, free path of beam)
-
Check log data for cause of error
-
Is signal via interface active?
-
Check for contamination of the sensor glasses on inside
and outside
Motor error Baumüller or EncoTRive
-
Check supply voltage
-
Check error codes on the controller
Summary of Contents for 4350127.347
Page 1: ...thicknessCONTROL MTS 8202 LLT C frame Non contact thickness measurement Operating Instructions...
Page 10: ...Page 10 Overview Technical Data Safety Instructions thicknessCONTROL MTS 8202 LLT 4350127 347...
Page 56: ...Page 56 Installation and Initial Operation thicknessCONTROL MTS 8202 LLT 4350127 347...
Page 64: ...Page 64 Maintenance Service Spare Parts Lists thicknessCONTROL MTS 8202 LLT 4350127 347...
Page 66: ...Page 66 Declaration of Incorporation thicknessCONTROL MTS 8202 LLT 4350127 347...