Technical specifications, calibration, and applicable standards
134 Baker DX 71-030 User Guide EN V10
www.megger.com
Table 5. Tests (all models).
Surge PD
Inception and extinction voltages (PDIV,
PDEV):
Measured per IEC 61934
Repetitive inception and extinction
voltages (RPIDV, RPDEV):
Measured per IEC 61934
Programmable PD threshold range:
1.0 mV – 999 mV
PD time resolution (per pixel):
10 nS – 50
μ
S
Resistance
Inductance
Source voltage,
maximum:
3.9 V
Source voltage,
maximum:
3.9 V
Source current:
600 mA
Source current,
maximum:
600 mA
100 to 10,000
Ω
:
3% accuracy
Source frequency:
50 to 4000 Hz
0.2 to 100
Ω
:
2% accuracy
1,000 to 5,000 mH
at 120 Hz:
15% accuracy
0.002 to 0.2
Ω
:
4%, ±1 m
Ω
100 to 1,000 mH
at 120 Hz:
8% accuracy
0.05 to 100 mH
at 1 kHz:
5% accuracy
160 to 5,000 mH
at 60 Hz (DX-15
only):
3% accuracy
0.5 to 160 mH at
60 Hz (DX-15
only):
2% accuracy
0.05 to 0.5 mH at
60 Hz (DX-15
only):
3% accuracy
Capacitance
Impedance
Source voltage,
maximum:
3.9 V
Source voltage,
maximum:
3.9 V
Source current,
maximum:
600 mA
Source current,
maximum:
600 mA
Source frequency:
4000 Hz
Source frequency:
50 to 4000 Hz
0.04 to 2.6
μ
F at
4000 Hz:
3% accuracy
0.15 to 10 000
Ω
at 60 Hz:
3% accuracy
2.6 to 26
μ
F at
4000 Hz:
5% accuracy
0.01 to 0.15
Ω
at
60 Hz:
3% accuracy
Phase accuracy at
60 Hz:
< 2 degrees
Summary of Contents for Baker DX
Page 1: ......
Page 2: ......
Page 10: ...Table of Contents viii Baker DX 71 030 User Guide EN V10 www megger com...
Page 28: ...Baker DX Instrument Overview 18 Baker DX 71 030 User Guide EN V10 www megger com...
Page 38: ...Baker DX User Interface Overview 28 Baker DX 71 030 User Guide EN V10 www megger com...
Page 58: ...Setting up the Baker DX Tester 48 Baker DX 71 030 User Guide EN V10 www megger com...
Page 149: ...megger com 139 RIC Templates Figure 141 RIC template 1...
Page 150: ...RIC templates 140 Baker DX 71 030 User Guide EN V10 www megger com Figure 142 RIC template 2...
Page 155: ...Index megger com 145...
Page 156: ...megger com 146...