Procedures for Testing with the DX Host
megger.com
95
Using the Recall Reference Feature
If you come in fresh to this screen, you have to create a reference. If you test a coil and touch
the
Set Reference
icon, that data becomes your reference. But if you have not tested anything
yet, and you already have a reference saved, touch the
Set Reference
icon and the following
dialog box appears so you can locate and select a previously-saved reference.
1.
Select folder, record, then reference. You will see a dialog confirming that the new
reference was set.
Figure 105. Using the recall reference feature.
2.
When you have set the new reference and before you start testing other coils, touch
the
ZS Override
icon to establish its settings at the reference level.
3.
Touch
Yes
when the following message appears:
Figure 106. Override zero-start functionality message dialog.
4.
After you set the ZS Override level, the
Voltage Output Control
knob on the unit will
no longer control the unit’s output voltage.
Summary of Contents for Baker DX
Page 1: ......
Page 2: ......
Page 10: ...Table of Contents viii Baker DX 71 030 User Guide EN V10 www megger com...
Page 28: ...Baker DX Instrument Overview 18 Baker DX 71 030 User Guide EN V10 www megger com...
Page 38: ...Baker DX User Interface Overview 28 Baker DX 71 030 User Guide EN V10 www megger com...
Page 58: ...Setting up the Baker DX Tester 48 Baker DX 71 030 User Guide EN V10 www megger com...
Page 149: ...megger com 139 RIC Templates Figure 141 RIC template 1...
Page 150: ...RIC templates 140 Baker DX 71 030 User Guide EN V10 www megger com Figure 142 RIC template 2...
Page 155: ...Index megger com 145...
Page 156: ...megger com 146...