| HCU 40 |
4 Operation
| 45 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
1
To close the failure report touch the
[Confirm]
symbol.
2
Take the HCU 40 out of service and have it tested by the authorized service
personnel.
4.2.2 Setting and Changing Setpoint Temperatures
NOTE
Preset temperature
After switching on the HCU 40, the setpoint temperature setting is always 37°C.
You can set the setpoint temperatures (T
set
) for the patient circuit and cardioplegia
water circuit independently of each other.
NOTE
Hotkeys
You can use hotkeys to call up the values for setpoint temperatures and setpoint
gradients directly instead of having to change them individually (⇨ "Changing the
Hotkey Settings", page 59).
NOTE
Maximum cooling capacity
If you set a setpoint temperature of 1°C, the internal mixing valve opens
completely to the tank. This provides the maximum available cooling capacity.
1
Touch the symbol "T
set
" for the circuit whose setpoint temperature you wish to
change.