4 Operation
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
NOTE
Recommendation for operating the HCU 40:
The HCU 40 should always remain connected to the power supply and turned
on to ensure that there is always sufficient ice in the tank. As soon as the set ice
block size is attained, the machine switches automatically to Ice maintenance
mode.
In Ice maintenance mode an increased quantity of ice is formed over a period of
several days without use of the HCU 40. To prevent the tank from freezing up
completely, the max. setpoint ice block size selected is automatically reduced to
"Medium ice block" (standard setting) from the 4th day in continuous Ice
maintenance mode (⇨ "Changing the Setpoint Ice Block Size", page 61).
Display following successful self-test
After the self-test, the touchscreen displays the main screen (⇨ "Main Screen",
page 28):
Main screen with external temperature sensors
Main screen without external temperature sensors
Display in case of error
Should an error occur during the self-test, the touchscreen displays a self-test
failure report.