2.1.4 Intended Patient
The device can be used for all patients irrespective of age, body weight and
gender.
2.1.5 Intended Environment
The HCU 40 is used in a clinical environment.
2.1.6 Contraindications
When the heater-cooler unit HCU 40 is used by specially trained personnel under
the supervision of a physician and in compliance with the intended use, no
contradictions are to be expected.
2.2 General Safety Instructions
2.2.1 Precautionary Measures
WARNING!
n
The system must be monitored by a trained member of specialist medical
staff. Clinical procedures and methods are the responsibility of the physician.
n
You should always keep a replacement unit on standby in order to ensure
continuous operation in the event of a complete system failure.
n
Prior to using the system, please read these Instructions for Use and the
Instructions for Use of all the disposables and supplies used.
n
The HCU 40 and all the system components used must comply with the
requirements of IEC 60601-1: 2005, section 16. In case of doubt, contact the
manufacturer of the system components used.
WARNING!
n
It is not permitted to change or modify the device or its accessories.
n
Switch the unit off and disconnect it from the external power supply before
maintenance, cleaning, or storage.
n
n
If a defective warming/cooling blanket is used with an electrosurgical unit
which is either not grounded or incorrectly grounded, this may result in burns
to the patient (⇨ "Connecting a Heat Exchanger", page 52).