4 Operation
| 39 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
4.1.2 Connecting a Control Unit
The CAN connection cable supplies the control unit with power and transfers data
between the HCU 40 and the control unit.
1
Connect the CAN connection cable to
the "CU"
[1]
connection on the HCU 40.
1
2
2
Connect the CAN connection cable to the "HCU"
[2]
connection on the CU
control unit.
4.1.3 Connecting External Devices (Optional)
WARNING!
n
If you are using the HCU 40 together with other medical devices, check the
total leakage current.
n
Only connect the device to connections that are intended to be connected to
other powered devices during normal operation.
n
Ensure that devices which are connected to the USB port of the HCU 40 fulfill
the specifications of the following standards:
‒
IEC 60950 (for data processing equipment located more than 1.5 meters
from the operating table)
‒
IEC 60601 (for data processing equipment located within 1.5 meters of
other medical devices)
n
Do not touch the patient and the device at the same time.
n
Only use the interfaces for the respectively intended devices.
Please observe the Instructions for Use of the external device for its setup and
operation.
The connections for external devices are located on the underside of the control
unit CU (⇨ "Control Unit CU", page 24).
4.1.4 Connecting External Temperature Sensors (Optional)
You can connect external temperature sensors to the HCU 40 in order to display
the measured temperatures (⇨ "Parameter Display", page 27) and to use the
gradient mode (⇨ "Gradient Mode", page 49).