| 144 |
12 Applied Standards
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment –
guidance
Electrostatic dis
charge (ESD)
IEC 6100042
±6 kV contact
±6 kV contact
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humid
ity should be at least 30%.
±8 kV air
±8 kV air
Electrical fast
transient/burst
IEC 6100044
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of
a typical commercial or hospital envi
ronment.
±1 kV for input/
output lines
±1 kV for input/out
put lines
Surges
IEC 6100045
±1 kV differential
mode
±1 kV differential
mode
Mains power quality should be that of
a typical commercial or hospital envi
ronment.
±2 kV common
mode
±2 kV common
mode
Voltage dips,
short interrup
tions and volt
age variations
on power supply
input lines
IEC 61000411
< 5% U
T
for ½ cy
cle (> 95% dip)
< 5% U
T
for ½ cycle
(> 95% dip)
Mains power quality should be that of
a typical commercial or hospital envi
ronment.
Thanks to its builtin battery, the de
vice continues to operate during power
mains interruptions. It therefore does
not need to be powered from an unin
terruptible power supply or external
battery.
40% U
T
for 5 cy
cles (60% dip)
40% U
T
for 5 cycles
(60% dip)
70% U
T
for 25 cy
cles (30% dip)
70% U
T
for 25 cy
cles (30% dip)
< 5% U
T
for 5 s
(> 95% dip)
< 5% U
T
for 5 s
(> 95% dip)
Magnetic field
with a power fre
quency
(50/60 Hz) as
per
IEC 6100048
3 A/m
3 A/m
The strength of powerfrequency mag
netic fields should correspond to that
of a typical commercial or hospital en
vironment.
Note:
U
T
is the AC supply voltage prior to application of the test level.
Portable and mobile RF communica
tions equipment should be used no
closer to any part of the device, includ
ing cables, than the recommended
separation distance calculated from
the equation applicable to the fre
quency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 6100046
3 V
eff
150 kHz to
80 MHz
10 V
eff
d
= 0.35 √P
Radiated RF
IEC 6100043
3↓V/m 80 MHz to
2.5 GHz
10 V/m
d
= 0.35 √P 80 MHz to 800 MHz
d
= 0.7 √P 800 MHz to 2.5 GHz