7 Cleaning, descaling and disinfection
| 105 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
7.2.2 Performing weekly routine disinfection
WARNING!
Use personal protection equipment (e.g. gloves resistant to chemicals made of
Nitrile rubber or Butyl caoutchouc (butyl rubber), dust mask for respiratory
protection P2, goggles and a protective lab coat) according to the Safety Data
Sheet for the disinfectant/descaling agent.
Disinfection is performed once a week with 2% Chloramine-T in the entire water
volume of the HCU 40. A cleaning wizard guides you through routine disinfection.
Total duration: approx. 150 to 200 minutes (depending on the amount of ice
present and the power supply).
During this procedure, the following steps are performed in succession:
n
Emptying of tank
n
Routine disinfection
n
Rinsing cycle 1
n
Rinsing cycle 2
n
Rinsing cycle 3
NOTE
The cleaning wizard provides additional support if one step in the cleaning
routine cannot be completed due to an operating error (example: shut-off valve
not open during pumping) A warning to this effect is shown. You can continue
the cleaning routine by pressing the
[Next]
symbol.
1
Touch the
[Settings]
symbol.
2
Touch the
[Cleaning/emptying]
symbol.
3
Touch the
[Cleaning]
symbol.
▶
You will see a wizard that will guide
you through disinfection.
4
Touch the
[Next]
symbol.