6 Messages
6.1 Alarms
6.1.1 Duration and Intervals for Acoustic Alarms
Higher priority alarms generate more alarm tones than alarms with lower priority.
Higher priority alarm tones are repeated at shorter intervals than alarms tones
with lower priority.
Priority
Alarm tones
Repeat interval
High
3+2 (c, d, e+f, g)
2.5 seconds
Medium
3 (c, d, e)
7.5 seconds
Low
2 (c, e)
20 seconds
6.2 Alarm List
The alarm list shows the last 6 alarms.
1
Touch the
[Settings]
symbol in the menu, and then touch
[Alarm history]
in the
"Settings" screen.
▶
The touchscreen displays the
following window:
1
2
3
1 Date
3 Message
2 Time
▶
▶
The HCU 40 displays current alarms, according to priority, on a colored
background: Red: high-priority alarm, Yellow: low or medium-priority
alarm.
2
To close the alarm list, touch the
[Confirm]
button.