2.2.7 Electromagnetic Compatibility
The HCU 40 complies with the requirements of the IEC 60601-1-2 standard on
electromagnetic compatibility. The system and all accessories and sensors fulfill
the EMC requirements of a typical clinical environment. (⇨ "Electromagnetic
Compatibility (EMC)", page 143)
The user is responsible for ensuring that the clinical environment complies with
the limits prescribed in IEC 60601-1-2. Exceeding these limits may impair the
system's efficiency and safety.
WARNING!
n
Do not use the HCU 40 in the vicinity of devices that emit high-frequency
signals (e.g., cell phones or high frequency devices). These can cause
excessively strong electromagnetic interference that exceeds the compliance
level of the HCU 40.
n
Observe normal precautions regarding relative humidity and the electrical
conductivity of clothing in order to minimize the build-up of electrostatic
charges.
n
To ensure safe use, the length of all connection cables of the HCU 40 must
not be changed.
n
Only use the specified accessories (⇨ "Accessories", page 133). The use of
other devices, systems, or accessories may increase RF emissions or reduce
the immunity to interference.
2.3 Symbols
Symbols on the rating plates of the HCU and control unit (CU)
Notice! Observe the warnings and safety precautions given in the accompanying docu
mentation.
Warning: Dangerous voltage
Observe the instructions in the Instructions for Use!
Protection type in accordance with IEC 60529: Protection against ingress of solid foreign
objects larger than 12.5 mm and dripping water when tilted up to 15°.