| 136 |
11 Technical Data
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Cooling system
Continuous cooling capacity of the compressor
4867 kJ/h (1352 W)
Heating system
Electrical heaters
Heating capacity
n
2 x 3000 W (200 … 240 V)
n
2 x 1500 W (110 … 120 V)
Power supply
Max. power consumption
n
1760 … 1920 VA (110 … 120 V, 16 A)
n
2760 … 2880 VA (230 … 240 V, 12 A)
n
2990 VA (230 V, 13 A)
n
3000 … 3600 VA (200 … 240 V, 15 A)
n
3200 … 3840 VA (200 … 240 V, 16 A)
AC power supply:
n
Mains voltage (set at the factory)
110/115/120/200/208/220/230/240 V
n
Frequency
50/60 Hz
n
Line fuse
25 A (110 … 240 V)
Supply line (Building installation requirements)
n
Exclusively used for HCU 40
n
Designed for current rating > 16 A (country
specific)
n
Protected by type C circuitbreaker type C
and ground fault circuit interrupter
Water supply
Water hardness
≤ 2.5 mmol/l CaCO3 (14 °dH)
Do not use deionized water!
11.2 Control Unit CU
Dimensions (H x W x D)
174 x 210 x 70 mm
Weight (incl. holder)
1.3 kg
Touchscreen
LCD, 115.2 x 86.4 mm, 640 x 480 pixels
11.3 Permissible Heat Exchangers
11.3.1 Oxygenator and Cardioplegia Heat Exchangers
Permissible water flow
Cardioplegia heat exchanger: at least 2 l/min
Oxygenator heat exchanger: at least 3 l/min
Summary of Contents for HCU 40
Page 1: ...INSTRUCTIONS FOR USE HEATER COOLER UNIT HCU 40...
Page 147: ......