4 Operation
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Ice size parameter
Meaning
Setpoint ice block size:
Medium ice block
A medium ice block is to be produced (standard setting).
Setpoint ice block size:
Large ice block
A large ice block is to be produced.
4.4.3 Changing the Water Flow
This function can be used to change the flows for the patient and cardioplegia
water circuit.
1
Touch the
[Settings]
symbol.
2
Touch the
[Water flow]
symbol.
3
To change the values, touch the
corresponding field and adjust the value
(⇨ "Changing Numerical Settings", page
35).
4
To accept the changes, touch the
[Confirm]
symbol.
Or
If you wish to reject the changes, touch the
[Reject]
symbol.
4.4.4 Changing the Settings for Locking the Controls
WARNING!
Risk from accidental changing of the settings.
When the locking of the controls is deactivated, there is a risk that settings may
be inadvertently changed, resulting in unsuitable values being accidentally set.
This function enables you to change the automatic locking of controls (⇨ "Locked
Controls", page 36). In addition, you can also deactivate the automatic locking.
1
Touch the
[Settings]
symbol.
2
Touch the
[System lock]
symbol.
3
Select from the list the period of inactivity after which the HCU 40 is locked
automatically.