6 Messages
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Message
Possible conse
quences
Possible cause
Possible action
Water level and ice
sensor defective!
Danger of frozen tank!
If necessary, stop com
pressor to reduce ice
thickness! Vis. check
water and ice level.
Tank frozen.
Water level and ice
sensor defective!
Vis. check water and
ice level. If necessary,
stop compressor to re
duce ice thickness! No
tify authorized service.
Keep/use replacement
device.
HLM connection faulty! Device cannot be con
trolled via heartlung
machine.
CAN communication
with HLM interrupted.
Check connection vis
ually. Notify the autho
rized service if neces
sary.
Ice monitoring defec
tive! Change device or
operate the compressor
manually. Check ice.
If ice formation is too
great: the water tank
may freeze. No circula
tion and thus no tem
perature regulation
possible.
Ice sensor defective.
Start the ice formation
manually if necessary.
Replace device as
quickly as possible. No
tify authorized service.
Use replacement de
vice.
Too little ice
Water level and ice
sensor defective!
Water level sensor de
fective! Check water
level visually
Displayed water level is
incorrect
Level sensors defec
tive.
Notify authorized ser
vice. Use replacement
device. Check water
level visually. Top up
water if necessary.
Top up water in tank
If water level falls fur
ther: No circulation with
empty tank, and there
fore no temperature
regulation possible.
Too little water in the
tank.
Fill the water tank.
Level sensors defec
tive.
If the error occurs
again: Notify authorized
service. Use replace
ment device.
Pat.: Tank temperature
too low. Danger of
frozen tank! Check ice
thickness visually. If
necessary, reduce ice
formation.
Tank may freeze.
Sensor faulty.
Reduce setpoint ice
block size. Check ice
thickness visually.