4 Operation
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Ice size parameter
Meaning
No ice
No ice should be formed.
Setpoint ice block size:
Small ice block
A small ice block is to be produced.
Setpoint ice block size:
Medium ice block
A medium ice block is to be produced.
Setpoint ice block size:
Large ice block
A large ice block is to be produced.
No ice buildup
The compressor is switched off, no ice demand.
Ice buildup active
The compressor is switched on, ice buildup taking place
until setpoint ice block size is reached.
Ice buildup completed
The compressor is switched off; the setpoint ice block size
has been reached.
4.5.3 Displaying the Pressure Limits
WARNING!
n
You must only use disposables (heat exchangers) which have a permissible
maximum pressure equal to or higher than the set pressure limit.
You can use this function to check the set pressure limits for the circuits.
1
Touch the
[Settings]
symbol.
2
Touch the
[System]
symbol.
3
Touch the
[Pressure limit]
symbol in the
system settings.
4
To close the window, touch the
[Confirm]
symbol.
4.5.4 Testing the Functioning of the Speaker and Warning Buzzer
WARNING!
Mortal danger to patient if acoustic signals do not sound!
n
If no acoustic signals sound, stop using the device and contact the authorized
service (⇨ "Authorized Service", page 131).