3 System Description
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Function
Meaning
Service
The
[Service]
function is passwordprotected and may only
be used by authorized personnel (
Brightness/volume
Changing the brightness/volume (
Time/date
Change the time, date and formats (
Time, Date and Formats", page 63)
Language
Pressure limitation
3.10 Pausing the Current Alarm
When alarms are paused, the alarm is only displayed optically. The acoustic
alarm pause lasts for a maximum of one minute.
The alarm pause can only be activated during an alarm. It only applies to the
current alarms. If a new alarm is triggered during the pause, the HCU 40 will
generate another acoustic alarm.
Alarm pause
Acoustic alarms enabled
To pause the current alarms, touch the symbol.
Pausing the current alarms
Acoustic alarms are generated again once the pause ex
pires or new alarm situations arise.
To end the alarm pause, touch the symbol.
Not possible
There is no current alarm that could be paused.
3.11 Basic Handling Information for Software
3.11.1 Confirming or Rejecting Inputs/Changes
If you enter or change data you must confirm or reject it. To this end, the
touchscreen displays the following symbols:
Symbol
Description
Confirm
To confirm the inputs or changes, touch the symbol.
The HCU 40 uses the new, changed settings.