| HCU 40 |
4 Operation
| 57 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
▶
P1: Oxygenator heat exchanger to water circuit P1 (⇨ "Connecting a Heat
Exchanger", page 52).
Or
▶
P1: Pressure reducer for warming/cooling blanket (⇨ "Connecting a Heat
Exchanger", page 52).
▶
P2: Connecting the warming/cooling blanket to the water circuit P2 (⇨ "
Connecting a Heat Exchanger", page 52).
Or
▶
CPLG (⇨ "Connecting a Heat Exchanger", page 52).
2
▶
The default temperature is set as 37°C.
3
Touch the
[Functions]
symbol in the
main screen.
4
Touch the
[De-airing]
symbol of the required circuit.
▶
The touchscreen shows the following
message:
5
To confirm the message, touch the
[Confirm]
symbol.
▶
The HCU 40 starts the pump to de-air the circuit.
▶
A status bar displays the progress of the de-airing.
NOTE: You can cancel the de-airing by touching the
[De-airing]
symbol
during the de-airing.
▶
Once the circuit has been de-aired, a green check mark appears next to
the
[De-airing]
symbol.
▶
The de-airing is followed by temperature regulation. The water is
regulated to the preset temperature.
NOTE: During de-airing, water is taken from the tank, and water flows into