| 140 |
11 Technical Data
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Option
Possible settings
a
Factory setting
n
Volume
1 ... 3 (in increments of 1)
3
Languages
b
German, English, Spanish,
French, Portuguese
English
Time/date:
n
Date format
DD.MM.YYYY, MM/DD/YYYY
DD.MM.YYYY
n
Time format
hh:mm:ss, hh:mm
hh:mm:ss
12 h, 24 h
24 h
a.
The possible values correspond to the setpoint temperatures and gradients.
b.
Changes remain when device is turned off and on.
Parameter
Factory setting "Cooling"
hotkey
"Warming" hotkey
Setpoint temperature:
n
Cardioplegia water circuit
a
4°C
37°C
n
Patient water circuit
a
34°C
37°C
Setpoint gradient:
n
Cardioplegia water circuit
Off
Off
n
Patient water circuit
Off
Off
a.
Changes remain when device is turned off and on.
11.8 Availability of Physiological Alarms for External Devices
No physiological alarms are provided for external devices.
11.9 Essential Performance Characteristics
The HCU 40's safety system monitors the temperature regulation. When a water
temperature of 42.0 ± 0.5°C is attained, the safety system switches off the pump
and heaters of the machine immediately, and a high-priority visual and acoustic
alarm is triggered. The pump can only be switched back on again once the water
temperature has fallen below 41.0°C.
Temperature overshoots can be a maximum of 0.5°C (± 0.5°C measuring
accuracy) above 41.0°C for 10 seconds. When this limit is exceeded, the heater
switches off immediately, and a high-priority alarm is emitted optically and
acoustically.
A water temperature of 42.0 ± 0.5°C can occur in a type 1 error situation for a
maximum of 60 seconds.