WARNING!
n
Observe the Instructions for Use for the heat exchangers employed.
n
Observe the permissible values for heat exchangers (⇨ "Permissible Heat
Exchangers", page 136).
n
The pressure limit must be set by the authorized service personnel in
accordance with the permissible pressure of the heat exchangers. Have the
pressure limit adjusted by the authorized service personnel if you use other
heat exchangers with a lower permissible maximum pressure than the set
pressure limit.
n
Use the HCU 40 and heat exchanger at the same height in order to avoid an
increase in pressure between the HCU 40 and the heat exchanger.
n
The water tank must only be filled with sterile filtered water and other
substances which have been specified by Maquet.
n
Check all water tubes and tube connections for leaks prior to the application.
n
Keep the tubes away from sources of heat.
n
The length of the tube from the HCU 40 to the heat exchanger must be at
least 1 m.
WARNING!
n
Turn the cardioplegia water circuit pump off if no cardioplegic solution is
required.
n
If the tubes are not connected to a heat exchanger, connect the ends of the
tubes to the cleaning connector.