4 Operation
| 53 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
3
Cardioplegia heat exchanger
P1
Patient water circuit 1
P2
Patient water circuit 2
CPLG
Cardioplegia water circuit
P
Ven
Venous pressure sensor
P
Art
Arterial pressure sensor
T
ext
External temperature sensor
Patient water circuit 1 (P1)
Patient water circuit 2 (P2)
Cardioplegia water circuit
(CPLG)
Oxygenator heat exchanger
–
–
Oxygenator heat exchanger
–
Cardioplegia heat exchanger
–
–
Cardioplegia heat exchanger
Oxygenator heat exchanger
Warming/cooling blanket
Cardioplegia heat exchanger
Pressure reducer
Warming/cooling blanket
Cardioplegia heat exchanger
Connecting the oxygenator heat exchanger to water circuit P1
WARNING!
n
Before use, check the pressure resistance of the heat exchanger, as shown
on the manufacturer's data sheet.
n
The heat exchanger used must be approved for a water flow of at least 2 l/
min.
WARNING!
n
The tubes of the perfusion system used should not be too long.
n
Insert the external temperature sensors (optional) fully into the measurement
device in the perfusion system and lock into place.
1
Connect the water tube from the water
outlet
[8]
to the water inlet of the
oxygenator heat exchanger.