| 128 |
8 Maintenance
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
8 Maintenance
Maintenance includes all measures which ensure the device is functioning
properly and is free of damage in order to enable safe use for the intended
purpose:
n
Maintenance by the operator
n
Inspection and maintenance by authorized service personnel
n
Repair
Interval
Task
Procedure
After each use
Surface cleaning and disinfec
tion of the device
Operator
Daily
Daily inspection by the operator Operator
Weekly
n
Perform routine disinfection
of the water circuits
n
Perform diagnosis
Operator
Monthly
Clean the air filter
Operator
Every 3 months
Perform descaling
Operator
Every 12 months
Inspection
Authorized service
8.1 Maintenance by the Operator
This section contains all of the regular inspection and maintenance measures
which are required to ensure the device is functioning properly and is free of
damage.
8.1.1 Daily Inspection by the Operator
Check the following points every day:
n
The water level in the tank is sufficient (setpoint water level: level marking in
the cardioplegia tank, approx. 1 cm above the evaporizer plates).
n
The size of the ice block is sufficient.
n
The interior walls of the tank must not be touched by the ice in order to allow
sufficient flow.