6 Messages
| 83 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
6.4.1 High Priority
Message
Possible conse
quences
Possible cause
Possible action
Cplg.: Heater tempera
ture too high!
Patient temperature not
regulated until device
cools down.
Temperature turned off
because temperature at
outlet too high.
Use replacement de
vice.
Operation in too hot en
vironment
Reduce temperature at
outlet to below 39.5°C:
allow device to cool
down. Place device in
cooler environment.
Supply cold air. Use re
placement device.
Temperature in device
too high
Temperature sensor
defective
Check plausibility of the
temperature display on
the CU. Notify the au
thorized service in case
of defects. Use replace
ment device.
Cplg.: Water tempera
ture too high
Outlet temperature
above warning limit
Check setpoint temper
ature and reduce if nec
essary. Message disap
pears automatically
when temperature
drops below warning
limit.
Pat.: Heater tempera
ture too high!
Patient temperature not
regulated until device
cools down.
Temperature turned off
because temperature at
outlet too high.
Operation in too hot en
vironment
Reduce temperature at
outlet to below 39.5°C:
allow device to cool
down. Place device in
cooler environment.
Supply cold air. Use re
placement device.
Temperature in device
too high (temperature
buildup in standby)
Temperature sensor
defective
Check plausibility of the
temperature display on
the CU. Notify the au
thorized service in case
of defects. Use replace
ment device.