Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
A-10
248083227-1 Rev 8
Data Record
Format
GENERAL
Each REMOTE counter can send a record of its data. The record is in
the form of an ASCII string. The meaning of a particular character is
dependent upon its position in the string.
Fields within the string are separated with spaces.
The first character in the response is the echoed command. This
character is not included in the Checksum value.
Figure A-1 shows the data format of a record for an eight channel
counter. Only two channels are shown for clarity.
Figure A-1 Data Record Serial Format
A$ 082603 131802 0100 0.1 001557 1.0 000293 LOC 000001 C/S 000BEA[CRLF]
Status
Date Time Interval Ch 1 (Ch 2-7) Ch 8 Location Checksum End Msg.
*CRLF means carriage return and line feed
Size Value Size Value Tag Value Tag Value
Command
...
Sample