Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
5-2
248083227-1 Rev 8
ASCII Command Structure
The commands are defined as case-sensitive ASCII characters. Format
is defined as:
Note:
The < and >
characters are part of the
command and must be
typed. The brackets [ and ]
are field delimiters and are
not typed.
<xx [yyyy]>
where:
<
= Start Character
Note:
The commands
are case sensitive. If an
incorrect case is used,
there will be no response
from the instrument.
xx
= Command Character
yyyy = Optional Command Parameter
>
= End Character
Handshake
Upon execution of the command, and assuming the user has typed
<V>
for Verbose, the REMOTE instrument will send a response based on
the command sent.
When the unit is powered up, it sets up the counters and timers, and
starts reporting data. The data will be similar to this example:
Note:
the field bk= 371
in this example, is the
"backlight" function, which
is a measurement of
scattered light in the
instrument.
371mV=.371V
Sample: 0, 0; bk= 371; trh= 74.7/ 43.0; rec=95;
Sample: 0, 0; bk= 371; trh= 74.7/ 43.0; rec=96;
If the unit is a 4-channel instrument, the following would display:
Sample: 0, 0, 0, 0; bk= 371; trh= 74.7/ 43.0; rec=95;
Sample: 0, 0, 0, 0; bk= 371; trh= 74.7/ 43.0; rec=96;
If an invalid command is typed, there will be no response.
Getting Started
1.
Start HyperTerminal with the above configuration settings.
2.
Connect the RLPC instrument with an RS-232 connection from
the 232/485 IN port to the PC COM port.
3.
Power on the RLPC.
The following menu will display:
LWS 3014 Remote/ ASCII.
Model: REMOTE LPC 0.2, LPC
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......