Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
B-4
248083227-1 Rev 8
Registers 40200-40999 are reserved for eight character names
associated with location index values. Thus the name for location =3
would be located at registers 40208-40211. Up to two hundred
locations can be specified.
Register 40199 indicates the number of location names supported on
this device.
40035
unsigned integer
Data Set [high]. Works in conjunction with 40036. Data entered
here is applied to the device through the command register.
40036
unsigned integer
Data Set [low]
40037
unsigned integer
Alarm Mode. Type of alarming performed
40038
unsigned integer
Alarm Parameter. Control parameter for given alarm mode.
40039
unsigned integer
Laser Reference Voltage (millivolts)
40040
unsigned integer
View Volume. Divide by 100 to get percentage.
For example: 6550d = 65.50%
40041
ASCII string
Flow Unit. Defines unit as cfm, lpm, mlpm char[0], char[1]
(NULL terminated string).
40042
ASCII string
Flow Unit. char[2], char[3]
40043
unsigned integer
Calibration Reference Voltage (millivolts)
...
40199
unsigned integer
Number of available alphanumeric location names (0 means
alphanumeric names are not supported).
40200
ASCII string
Location_1_char[0], char[1] (NULL terminated string)
40201
ASCII string
Location_1_char[2], char[3]
40202
ASCII string
Location_1_char[4], char[5]
40203
ASCII string
Location_1_char[6], char[7]
...
40996
ASCII string
Location_200_char[0], char[1] (NULL terminated string)
40997
ASCII string
Location_200_char[2], char[3]
40998
ASCII string
Location_200_char[4], char[5]
40999
ASCII string
Location_200_char[6], char[7]
Table B-3 Sensor Settings Registers
Register Data
Type
Description
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......