Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
6-8
248083227-1 Rev 8
Set Date & Time(WD)
Note:
There must be no
spaces between
characters and leading
zeros must be used.
This function is used to set the Real-Time Clock. This clock is used to
date/time stamp each data record when stored. Data format is month
day year hours minutes seconds.
There are no settings for time zones or Daylight Savings time.
Command:
WDmmddyyhhmmss[CRLF]
{No spaces}
Response:
WDmmddyyhhmmss[crlf]
Query by typing:
WD[CRLF]
Response:
WDmmddyy[space]hhmmss[crlf]
Counter
Setup and
Data Record
Commands
This group of commands is divided into three functional sections:
counter setup commands, buffered record (data) commands and
miscellaneous commands.
The general syntax form for each of these commands is a single capital
letter.
Counter Set Up Commands
These case-sensitive commands are used to set up the sample and hold
intervals for the instrument. No spaces are allowed within the
commands.
Set Hold Interval (H
hhmmss
)
This command sets or queries the instrument’s Hold Interval.
Example Command:
H001000[CRLF]
{Set Hold Interval to 10 minutes}
Leading zeroes can be left out:
H2000[CRLF]
{= 20 minutes}
H100[CRLF]
{= 1 minute}
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......