Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
A-4
248083227-1 Rev 8
If an invalid value is sent, the device will respond with a single
?
WD[ENTER]
WDmmddyy hhmmss[CRLF]
Displays current Date and Time.
There are no settings for time
zones or Daylight Savings Time.
H[ENTER]
Hhhmmss[CRLF]
Displays current HOLD
INTERVAL in hours, minutes,
seconds
L[ENTER]
Lhhmmss[CRLF]
Displays current SAMPLE
INTERVAL in hours, minutes,
seconds
Table A-2 Command Set for Displaying Current Configuration
COMMAND
RESPONSE
DESCRIPTION
Table A-3 Command Set for Configuration Parameters
COMMAND
RESPONSE
DESCRIPTION
WAd[ENTER]
WAd[CRLF]
Sets Alarm Limit to
value
d
which is a
numeric value from 1 to
999999.
WLd[ENTER]
WLd[CRLF]
Sets the software
configurable Location
Address of the device to
value
d
which is a
numeric value from 0 to
63.
WDmmddyyhhmmss[ENTER]
WDmmddyyhhmmss[CRLF]
Sets Real-Time clock
using format: month day
year hour minute second
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......