MR Protocol Commands
248083227-1 Rev 8
A-5
Counter
Setup
Commands
Table A-4 displays the commands that are used to setup the sample and
hold intervals for the instrument.
For the Sample and Hold Intervals, 1500=15 minutes, 100=1 minute,
60=1 minute.
Action
Commands
Action commands are used to set the counting mode and to control the
starting and stopping of the counter. See Table A-5.
The general syntax for Action commands is a single lowercase letter
with no arguments. The responses for the Action commands do not
have a carriage return/line feed [CRLF].
Table A-4 Particle Counter Setup Commands
COMMAND
RESPONSE
DESCRIPTION
L[ENTER]
L
hhmmss
[CRLF]
Queries the instrument’s
Sample Interval; displayed in
hours minutes seconds.
L
hhmmss
[ENTER]
L
Sets the Sample Interval to
value
hhmmss
. Leading zeroes
can be left out.
H[ENTER]
H
hhmmss
[CRLF]
Queries the instrument’s Hold
Interval; displayed in hours
minutes seconds.
H
hhmmss
[ENTER]
H
Sets the Hold Interval to value
hhmmss
. Leading zeroes can be
left out.
Table A-5 Command Set for Action Commands
COMMAND
RESPONSE
DESCRIPTION
a
a
Auto Mode:
Puts the counter into
Auto Mode. When used with the
[d]
Start command, unit will
continue to sample for the
specified Sample Interval and
Hold Interval repeatedly until a
Stop
[e]
is issued or manual
mode is selected.
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......