Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
4-2
248083227-1 Rev 8
DIP SWITCH GENERAL DEFINITIONS
Table 4-1 displays the general DIP Switch settings. OFF (DOWN) = 0,
ON (UP) = 1
Note:
Use a tool with a
very small, pointed tip in
order to change the DIP
Switch positions.
Table 4-1 DIP Switch settings
Communication Modes
There are three communication modes for the RLPC - MODBUS
protocol, MR protocol and ASCII mode as shown in Table 4-2.
The DIP Switches must be set before the unit is powered up. If the DIP
Switches are changed, power cycle the instrument to implement the
change.
Position
#
Description
Setting
1
Binary Bit 0
Addressing, OFF=0, ON=1
2
Binary Bit 1
Addressing, OFF=0, ON=1
3
Binary Bit 2
Addressing, OFF=0, ON=1
4
Binary Bit 3
Addressing, OFF=0, ON=1
5
Binary Bit 4
Addressing, OFF=0, ON=1
6
Binary Bit 5
Addressing, OFF=0, ON=1
7
Communication
Mode
Used in conjunction with DIP
switch 8 to set MR, MODBUS or
ASCII mode
8
Communication
Mode
Used in conjunction with DIP
switch 7 to set MR, MODBUS or
ASCII mode
Table 4-2 DIP Switch configuration for Communications
Mode
COMMUNICATIONS MODE
DIP
SW 7
DIP
SW 8
MODBUS protocol
OFF
OFF
ASCII mode
OFF
ON
MR Protocol with software addressing
ON
OFF
MR Protocol with DIP Switch addressing
ON
ON
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......