Configuring with LWS ASCII Service Protocol
248083227-1 Rev 8
5-13
Example: Setting Alarm Threshold <axy>
Where x is the channel number and y is the threshold value
.
The REMOTE instrument is preconfigured at the factory with an alarm
threshold of 1000 counts. This means that the counter will enable the
external alarm when the threshold of the alarm channel configured
exceeds 1000 counts. To view the currently configured alarm threshold
on the instrument,
1.
Type:
<?>
This command displays the current alarm threshold within
the parameter list.
To change the alarm threshold for Channel 1:
2.
Type:
<a1100>
Response:
Ch1 Alarm Threshold = 100 counts
Ch2 Alarm Threshold = 1000 counts
Ch3 Alarm Threshold = 1000 counts
Ch4 Alarm Threshold = 1000 counts
Alarm Thresholds only apply when alarm channels are enabled.
3.
To change the alarm threshold for channel 3 type:
<a3500>
Response:
Ch1 Alarm Threshold = 100 counts
Ch2 Alarm Threshold = 1000 counts
Ch3 Alarm Threshold = 500 counts
Ch4 Alarm Threshold = 1000 counts
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......