Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
6-12
248083227-1 Rev 8
Universal
Actions
Commands
General
These commands are provided for controlling multiple counters on the
same bus. The commands are not echoed.
WARNING:
The MR
Protocol is case
sensitive. Using the
incorrect case will
produce undesirable
results.
As soon as a Universal Action Command is executed, the bus drops out
of Universal mode.
Command:
ua
{
Universal Auto Sample Mode
-
Puts the counters in Auto
Count mode. When the
ud
command is used, the devices
will start counting in the auto mode, continuously executing
individual
Sample and Hold Intervals until a stop
command
ue
is sent
.
}
ub
{
Universal Manual Sample Mode
-
Places the counters in
Manual
C
ount mode. When the
ud
command is sent, the
devices will count in the Manual mode, each device
counting through its Sample Interval one time.}
uC
{
Universal Clear Buffer -
The contents of all buffers are
erased.}
uc
{
Universal Start Count (Computer Controlled)
-
The
instruments start counting and continue to count until a
stop command is sent from the computer.}
ud
{
Universal Start Count (Counter Controlled) -
The
instruments start counting according to each unit
’
s Sample
Interval and auto/manual presets.}
Note:
The last data
record will show 0000 for
Sample Interval, as the
Sample Interval is
controlled by the host.
ue
{
Universal Stop Count -
The instruments stop counting
immediately
and each instrument builds a data record.}
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......