Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
3-6
248083227-1 Rev 8
Inlet Nozzle
Note:
The user is
responsible for
maintaining a nominal
flow rate of 100 ml/min. +
5% through the sensor.
The sample source is attached using ¼" flare tubing to the Inlet Nozzle.
The user is responsible for maintaining a nominal flow rate of 100 ml/
min. + 5% through the sensor to insure the accuracy of the data. See
Figure 3-5.
Figure 3-5 Example of an RLPC Installation
If there are any questions or problems during system setup, please
contact Lighthouse Worldwide Solutions’ Technical Support at
[email protected].
Attaching
Sensor
This section instructs on how to attach and start using the RLPC. Refer
to previous sections of this chapter for electrical connection
information.
WARNING:
Please
follow the standard
precautions for working
with liquids around
electronic components.
Use of safety measures when working with liquids around electronics
cannot be stressed enough. Make sure the RLPC is physically secure
before connecting any tubing or electrical cables. Use Personal
Protection Equipment, such as face shields, protective garments and
gloves, whenever working around hazardous liquids.
Sample Source
Valve
Valve
Liquid Flow
Meter
Pump
(Optional)
Sample Return
or to Drain
Inlet
Nozzle
Outlet
Nozzle
RLPC
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......