<ta#>
Set Sample Seconds
Configures the sample time interval where # is the
number of seconds in the sample. The minimum value is
30 seconds and maximum value is 23 hours, 59 minutes,
59 seconds, or 83699 seconds.
<tb#>
Set Holdtime Seconds
Configures the time in between samples when the unit is
not counting; # is the hold time in seconds.
<pa#>
Set Location Num (0-999)
Configures the location of the instrument.
<aa#>
Set Alarm Channel #
Configures the instrument to use any channel to trigger
the alarm condition. Multiple channels may be used for
alarming.
<a1#>
<a2#>
<a3#>
<a4#>
Set Alarm # Threshold (# of counts)
Configures the value at which the alarm channel must
exceed in order to trigger the alarm condition. Each
channel may be configured to a different threshold.
<ae#>
Set Alarm Suppression #
The number of alarm conditions have to be greater than #
in order to turn on the Alarm Relay. Used for Local
Alarming.
<da#>
Set Year = YY
Configures the 2-digit year; ex: <da04> for 2004
<db#>
Set Month = MM
Configures the 2-digit month; ex: <db01> for 01
<dc#>
Set Day = DD
Configures the 2-digit day; ex: <dc28> for 28
<dd#>
Set Hour = HH
Configures the 2-digit hour; ex: <dd13> for
13
:00:00
<de#>
Set Minute = MMin
Configures the 2-digit minute; ex: <de25> for 13:
25
:00
<df#>
Set Second = SS
Configures the 2-digit second; ex: <df43> for 13:25:
43
Table 5-1 User Menu Commands <m1>
ASCII
Command
Description
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......