Configuring with LWS ASCII Service Protocol
248083227-1 Rev 8
5-3
Version: 110
Serial#: 040116001
2004-01-13, 18:05:26
- program mode:
4.
Type:
<V>
This is Verbose mode.
This must be the first command
sent to the instrument.
Response:
<V> Verbose Mode On
5.
Type:
<?>
This command displays the instrument’s current
configuration.
Response:
<?>
2004-01-24, 10:09:00
<Model: REMOTE LPC 0.2, LPC
LWS Protocol
View volume = 80.0%
Flow Rate = 100.0 MLPM
Sample Time = 60 secs
Hold Time = 3 secs
Service High = 3500 mV
Service Low = 125 mV
Location = 0
Device Address = 1
Ch1 Alarm = OFF
Ch2 Alarm = OFF
Ch3 Alarm = OFF
Ch4 Alarm = OFF
Ch1 Alarm Threshold = 1000 counts
Ch2 Alarm Threshold = 1000 counts
Ch3 Alarm Threshold = 1000 counts
Ch4 Alarm Threshold = 1000 counts
Suppress Alarms = 0
Thresholds (mV) = 40, 200
Channel 1 = 0.3
Channel 2 = 0.5