Programming with MR Protocol
248083227-1 Rev 8
6-9
H60[CRLF]
{= 1 minute}
H10[CRLF]
{= 10 seconds}
These are all valid commands.
Response:
H
Query by typing:
H[CRLF]
Response:
H001000[CRLF]
{Hold Interval = 10 minutes}
Set Sample Interval (Lhhmmss)
WARNING:
Do NOT
set the Sample Time
value to less than 30
seconds.
Sets or queries the instrument’s Sample Interval. The instrument is
programmed at the factory with a default sample time of one minute.
Do NOT set this value to less than 30 seconds.
Example Command:
L001500[CRLF]
{Set Sample Interval to 15 minutes}
Leading zeroes can be left out:
L1500[CRLF]
{= 15 minutes}
L100[CRLF]
{= 1 minute}
L60[CRLF]
{= 1 minute}
These are all valid commands.
Response:
L
Query by typing:
L[CRLF]
Response:
L001500
[CRLF]
{Current Sample Interval is 15 minutes}