Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
4-8
248083227-1 Rev 8
Make sure that each RLPC has a unique address (non-zero) or a
conflict will result and data will be lost.
The RLPC complies with EIA’s RS485 standards for distances and
number of devices on a chain. See Table 4-7.
Table 4-7 EIA Industry Standards for RS485 Communications
SPECIFICATIONS
RS485
Mode of Operation
Differential
Total Number of Drivers and Receivers on
One Line (One driver active at a time for
RS485 networks)
32 Drivers
32 Receivers
Maximum Cable Length
4000 ft. (1,219.2 m)
Maximum Data Rate (40 ft. - 4000 ft. for
RS422/RS485)
10Mb/s - 100Kb/s
Maximum Driver Output Voltage
-7V to +12V
Driver Output Signal Level (Loaded
Min.): LOADED
+/-1.5V
Driver Output Signal Level (Loaded
Max.): UNLOADED
+/-6V
Driver Load Impedance (Ohms)
54
Max Driver Current in High Z State
(POWER ON)
+/-100
μ
A
Max Driver Current in High Z State
(POWER OFF)
+/-100
μ
A
Receiver Input Voltage Range
-7V to +12V
Receiver Input Sensitivity
+/-200mV
Receiver Input Resistance (Ohms), (1
Standard Load for RS485)
>12k
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......