MR Protocol Commands
248083227-1 Rev 8
A-7
Note:
When the
ue
command is given, the last
data record will show 0000
for the Sample Interval as
the Sample Interval is
controlled by the host.
ub
Universal Manual Sample Mode
Puts the counters in Manual Count mode. When
the
ud
command is sent, the devices will count
through its Sample Interval and then stop. This
command is not echoed.
uC
Universal Clear Buffer
This command will erase the buffers on all the
devices. This command is not echoed.
uc
Universal Immediate Start
(Computer Controlled)
The devices start counting and continue to count
until a Stop command is sent from the computer;
data record does not record a sample time. This
command is not echoed.
ud
Universal Start Count
The devices start counting according to each
unit’s Sample and auto/manual presets. This
command is not echoed.
ue
Universal Stop Count
The instruments stop counting immediately with
this command and each instrument builds a data
record. This command is not echoed.
Table A-6 Universal Actions Command Set
COMMAND
DESCRIPTION
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......