21
Leica ST5020
Connecting the waste water hose
• Outlet connection piece (
10
) can be turned
to point into different directions, depending
on whether the drain water pipe is located
to the left or to the right of the instrument.
• To adjust the connection piece, loosen
screw (
11
) of hose clamp (
12
) sufficiently so
that the connection piece (
10
) can be tur-
ned. Adjust the connection piece as desired
(always with the opening pointing slightly
downwards) and retighten screw (
11
).
• Turn screw (
15
) to the left (use screwdriver)
until hose clamp (
14
) opens wide enough to
be slid over waste water hose (
13
).
• Fit waste water hose (
13
) onto outlet con-
nection piece (
10
) and position hose clamp
(
14
) as shown in
Fig. 9
.
• Turn screw (
15
) to the right until hose (
13
) is
secured tightly on connection piece (
10
) by
hose clamp (
12
).
• Insert waste water hose (
13
) into drain pipe
and secure in position.
Fig. 9
7
13
14
15
4. Setting up the instrument
Important!
When installing the drain hose al-
ways make sure there is a sufficient
descending gradient from the outlet
(connection piece (10)) on the instru-
ment to the drain pipe!