62
Instructions for Use, V2.0, RevG - 07/2016
Fig. 56
For a complete definition of a program step, the following parameters
have to be entered:
1. Station in which the step is supposed to be carried out.
2. Reagent corresponding to that step (must be allocated to the station).
3. Duration of the step.
4. Step duration must be defined as exact / non-exact.
5. Dips (yes/no).
6. Oven temperature (optional)
5.6.2 Defining the program steps
• Press
DETAILS
(
Fig. 54
) to access the
PROGRAMMING
menu (
Fig. 56
).
If a program is created from scratch, the list of the individual program
steps is completely empty.
• The number in the
STEP
column indicates the order in which the sta-
tions will be used.
Press
INSERT ROW
- a new program row (step) is created in the row
below and highlighted in blue/red.
• The total number of steps (new
programs automatically start from
number "
1
") is automatically in-
creased by
1
.
The
DURATION
,
EXACT
and
DIP
columns already contain standard
settings (
2
) which must be entered
depending on the program step.
• Code and name of the program (
1
)
are located above the table con-
taining the program steps.
• Program steps are displayed in the
order in which they are supposed
to be carried out.
• Up to 40 steps can be defined for
each individual program.
1
↓
2
5. Operation