20
Instructions for Use, V2.0, RevG - 07/2016
Fig. 8
Instrument rear panel
water in-/outlets
4. Setting up the instrument
10
11
7
12
Fig. 7
8
9
10
Instrument rear panel
- tap water inlets
• Screw tap water inlet hose (
7
in
Fig. 8
) for
tap water stations 8 - 12 onto water tap and
insert the other end into inlet (
8
in
Fig. 7
).
• If station no.
7
(
Fig. 3
) will also be used as
a tap water station, connect a second tap
water inlet hose (optional accessory) to in-
let (
9
).
Inlet (9) only provides water (or any
other medium supplied via a separa-
te connection) to station no. 7. To use
station 7 as a water station, a wash
vessel must be inserted and station 7
must be defined as a 'tap water' or
'fully-de-ionized water' station (DI
water station) in 'stations options'.
(see also
chapter 5.5.3
)